KE 1082 General purpose thermocouple with cable

KE 1082 General purpose thermocouple with cable

KE 1082
true
Sheath
Construction
Closed-end tube
Material
SS 316 - (Wst 1.4401) Also available in SS310, Inconel 600, Hasteloy X and Nicrosil D
Diameter
6.0mm standard, other diameters possible
Length
100mm standard, other lengths possible
true
Lead wire
Transition
Moisture resistant cable seal
Ingress protection
IP65 / IEC529
Thermocouple wire
PVC, PUR, PFA, FEP or specified by customer, thermocouple cable
Cable length
Customer to specify
true
Thermocouple
Calibration
K, T, J, E, N, U or L
For duplex
KK, TT, JJ, EE, NN, UU or LL
Accuracy
According IEC 60584
Hot junction
Insulated
Insulation test
Tested with 500 Vdc @ 20 °C
Insulation resistance
Minimum 100 M ohm
true
Options
Length
Other length possible
Diameter
Other diameter possible
Cable type
Other cable type possible

Customisation possibilities

You can remark the customisation in your quote

PRODUCTCODE

KE 1082